| Model | ATEKO, Radon reduction system (2015) |
|---|---|
| Operating time | 09:00 ~ 18:00 |
| Location | YangYang A5 tunnel (facility room) |
| inquiry |
1. Equipment name
- Radon reduction system (라돈 감쇄 장치)
2. Manufacture and Model
- ATEKO, Radon reduction system (2015)
3. Purpose
- To prevent the surface contamination by the Radon, This system supplies the air ,which the Radon is reduced by minimum 1,000 times, to the laboratories.
4. Specification and performance
- air filter, air drier, carbon columns, heat exchanger, charcoal filter, HEPA filter
- Medium: Air
. Supply (Radon reduced air) : 120 - 150 m3/h (ambient pressure, 20°C)
. Input radon concentration: 20 - 100 bq/m3
. Radon reduction factor: min. 1000
5. Location and Picture
- YangYang A5 tunnel (facility room)