IBS Institute for Basic Science
Search

Nitric Oxide Analyzer

equipment explanation
Model NOA 280i(GE Analytical Instruments)
Operating time 09:00 ~ 18:00
Location POSTECH Chem. Bldg. # 324
inquiry Prof. Kim, Wong Jong
054-279-9960
wjkim@ibs.re.kr

Explanation
Nitiric Oxide Analyzer (NOA) detects for nitric oxide, nitrite, nitrate or nitrosothiols in most biological samples. It measures liquid and exhaled breath nitric oxide with highly sensitive ozone-chemiluminescence technology. Addi tionally, Concentrations, ranging from low nanomolar to millimolar levels, can be measured in sample volumes ranging from a few microliters to several milliliters.

 

Specification
● Sensitivity
- Gas: < 1 ppb, Liquid: ~1 picomole
● Range
- Gas: <1–500,000 ppb, Liquid: Nanomolar to millimolar
● Response Time
- Electronics: 67 milliseconds to 90% full scale,Lag Time: 1 second
● Ambient Temperature- 32° to 86°F (0° to 30°C)
● Sample Size- Gas: 10–300 mL/min, Liquid: 0.001–5 mL
● Vacuum Pump with Installed Trap
- Height: 14.5 in (37 cm)
- Width: 7.5 in (19 cm)
- Length: 19 in (48 cm)
- Weight: 47 lbs (21.5 kg)
● Main Fuse
- 100 VAC model: T, 10A, 250V
- 120 VAC model: T,5A, 250V,
- 230 VAC model: T,5A, 250V
● Ozone Fuse
- 100 VAC model: T,200mA, 250V
- 120 VAC model: T,200mA, 250V